IntroductionHigh energy silver (AgL We have now for the first time implemented a silver monochromator as a modification to the standard aluminum monochromator on a modern magnetic lens based electron spectrometer, the Axis ULTRA. The anode of the aluminum monochromator has been adapted to have both a conventional Al (metal) face and an additional Ag face, thus a changeover from one source to the other may be made with only minor adjustment and without breaking vacuum. The great improvement gained in photoelectron collection efficiency by employing magnetic lens technology improves the sensitivity and ultimate usability of the silver source making it a viable alternative for the measurement of deep 1s core levels and Auger parameters. The results presented here show that the silver source gives a resolution better than 0.9 eV FWHM for a silver sample and a sensitivity greater than 4,300 CPS on gold at a resolution of 1.3eV. |
|||||||||||||||||||
Table 1: Higher energy X-ray sources
|
TheoreticalConventional commercial spectrometers are limited to
a maximum ionisation energy of 1486.6eV, the photon energy of the AlK Ag L
n
. |
||||||||||||||||||
Figure 1: X-Ray Monochromator Geometry
|
ImplementationIn practical terms it is necessary to run a modern photoelectron spectrometer
with better than 90% of uptime. To switch between aluminum and silver
monochromated sources it is unacceptable to vent the system and so lose
valuable instrument time. Therefore we have developed a twin anode monochromatic
source. The anode is provided with both aluminum and silver coatings
and two electron sources are used to generate Al or Ag X-rays as required
- see Figure 1.r |
||||||||||||||||||
| Figure 2: Gold survey spectrum using Ag mono |
ResultsPerformance
Gold foil (cleaned, survey spectrum: Fig. 2).
|
||||||||||||||||||
Figure 3: Al 2p regions: Al mono (a), and Ag mono (b) sources |
Greater analysis depthThe Al 2p spectra shown (Figure 3) were acquired with (a) Al and (b) Ag sources. Spectrum (b) is a clear illustration of the greater analysis depth of the Ag source as the element : oxide ratio is greater than that acquired with the Al source. An oxide thickness of 1.8nm indicated here was confirmed by independent angle resolved measurements. |
||||||||||||||||||
Figure 4: Si 2p (a) and Si 1s (b) regions, using Ag mono |
More information with higher photon energyDepth profiling can be performed by analysing photoelectrons of different binding energies. In this case the Ag source is used to analyse the Si 2p and Si 1s core levels from a thin native oxide on Si. Clearly the higher BE (lower KE) Si 1s spectrum has a greater oxide ratio (spectrum (b)) due to the reduced analysis depth when compared to the lower BE (higher KE) Si 2p spectrum, (a). |
||||||||||||||||||
Figure 5: Si 1s and Si KLL Auger series from Ag mono source. |
Higher energy Auger series and core levelsThe higher photon energy of the Ag source may be used to generate higher energy series. In this case Si KLL and Si 1s spectra were acquired from a native oxide on Si. |
||||||||||||||||||
ConclusionsThe silver monochromatic source implemented within the geometry constraints of a classical aluminum monochromator has been shown to offer viable, narrow line, high energy excitation. The unique combination of the two sources in the same physical unit means that the silver source is also a practical option for routine high energy analysis. In addition, improved performance of modern magnetic lens instruments means that although the Bragg equation (for silver) is only satisfied by a second order diffraction, useable count rates and realistic acquisition times are achieved. It is expected that the introduction of the dual anode monochromatic source will lead to more studies of higher energy core levels and non-destructive depth profiling. |
|||||||||||||||||||
References
|
|||||||||||||||||||